When sports chiropractors first appeared at the Olympic Games in the 1980s, it was alongside individual athletes who had experienced the benefits of chiropractic care in their training and recovery processes at home. Fast forward to Paris 2024, where chiropractic care was available in the polyclinic for all athletes, and the attitude has now evolved to recognize that “every athlete deserves access to sports chiropractic."
Technical Tips We May Have Forgotten
There are many reasons why x-ray films may lack the quality you expect. The least likely reason is the x-ray machine. Most of the time, it is the fault of the "technician," and we all know who that is.
You should treat your x-ray unit and dark room like a radiology department. We all have front desk procedures, reporting procedures, and billing procedures, so why not radiography procedures? The key to obtaining quality x-rays is consistency. Once good procedures are established, it only requires that those procedures are followed every time an x-ray is taken.
Here are a few useful recommendations:
- Have the calibration checked on your x-ray unit periodically (every 2-3 years).
- Warm up the x-ray tube before taking films. This will extend the life of your tube.
- Have a regular schedule for cleaning your cassettes (every 2-3 months for most offices). The busier the office, the more frequent the cleaning. Commercial cleaners are best, but make sure you don't scratch the screens.
- Chemistry in the processor should be changed at the very least every month. The more films taken, the more frequently it should be changed. The chemicals do degrade, so don't think you can just change them whenever.
- Clean the roller in the processor at least daily. Don't use anything abrasive.
- If you have an older processor, "feed" it films daily to keep the chemistry renewed.
- If you are still using tanks, change the fresh water every day. The developer and fixer should be clear. Don't wait until they're black to change them.
- Check your darkroom for light leaks at least once a year. The caulking dries up and ceiling tiles can shift.
- Keep a log with the patient's measurements and techniques used. If this begins to change, it can indicate a problem with the x-ray unit.
- For lateral lumbar x-rays, take two measurements: one at the waist and one at the hips, particularly with females. The median between the two should be the measurement used to calculate the technical factors. If the difference between the two measurements is greater than 6cm, plan on taking a lateral L5-S1 spot.
11. Remember to move the tube closer to the bucky one inch for every five degrees of tube tilt to maintain the original film focal distance. (I bet some of us have forgotten that one).
12. Identify the film. Proper identification includes:
- patient's name
- patient's age/sex, or date of birth
- exam date
- exam location
- right or left marker
These are the techniques I use for spinal views:
View | Distance | kVp | Angle | Central Ray |
C-Spine | ||||
APOM | 40" | 70 | none | uvula |
AP | 40" | 70 | 15 up | C4 |
Lateral | 72" | 70 | none | C4 |
Flex/extend | 72" | 70 | none | C4 |
Pillar | 40" | 70 | 35 down | C4 |
Oblique | 72" | 70 | 15 up or down | C4 |
Swim Lateral | 40" | 80 | 5 down | C7 |
T-Spine | ||||
AP | 40" | 80 | none | T6 |
Lateral | 40" | 80 | none | T6 |
L-Spine | ||||
AP | 40" | 80 | none | L3-4 |
Lateral | 40" | 90+ | none | L3-4 |
Oblique | 40" | 80+ | none | 2" above crest and 2" below medial ASIS |
AP L5-S1 | 40" | 80 | 20-30 up | between pubes and ASIS |
Lateral L5-S1 | 40" | 90+ | none | 2" below crest |
Sacrum | ||||
AP | 40" | 80 | 15 up | between pubes and ASIS |
Lateral | 40" | 90+ | none | through ASIS |
Sacroiliac | ||||
AP | 40" | 90 | 35 up | between pubes and ASIS |
Lateral | 40" | 90+ | none | same |
Coccyx | ||||
AP | 40" | 80 | 10 down | 2" above pubes |
Lateral | 40" | 80 | none | 4" below crest |
For more positioning techniques, refer to the Atlas of Radiographic Positioning by Sharon Jaeger,DC,DACBR.